Removing/Changing Router
The router may be changed or removed. Turn off the Patient
Simulator before doing so.
Open the pelvis and locate the router. Remove the two Ethernet
cables and black power cable.
If you would like to operate the Patient Simulator without a router,
then attach a through-adapter between the two Ethernet cables.
If you would like to install a new router, attach the power cable to
the DC plug and the Ethernet cables back into the router again.
Opening the Torso
Open the Patient Simulator torso for the following procedures:
Attaching or Replacing Limbs
− Attaching or dismantling the Patient Simulator legs and arms.
− Exchanging default arms for optional IV or trauma arms.
Maintenance Tasks
− Changing the Patient Simulator batteries.
− Replacing the Pneumothorax Bladders, Chest-Rise Bladders,
Lung Bladders, IO Modules and Chest Drain Modules.
− Replacing the Torso Skin.
− Performing a general inspection.
− Removing WLAN adapter.
To open the Torso Skin
1 Unzip the zippers on the Patient Simulator’s left shoulder and
torso.
2 Remove the Genitalia Module and release the skin flap from the
pelvis.
3 Fold the Torso Skin over to one side.
4 Open the Stomach Foam to one side, taking care not to tug on
the connecting tubes and cables.
Note: DO NOT disconnect the tubes and cables connecting the
Stomach Foam to the Patient Simulator.
5 Replace the Stomach Foam and close the Torso Skin, perform
steps 1- 4 in reverse.